1. Cover Plate
2. Nut and Washer
3. Eye Spacing Control
4. Viewing Head
5. Yaw Spigot
6. Height Friction Clamp
7. Low Voltage Socket
8. Anti Glare Shield
9. Objective/Selection Lever
10. x2 x4 Objectives
11. Illumination Adjustment Control
12. Low Voltage Plug
13. Supply Cable
14. On-Off Switch
15. Counter Balanced Stand
16. Supply Inlet Voltage Selector
17. Voltage Selector
18. Universal Mount (not shown)
19. Column Base
20. C Clamp (not shown)
21. Lighting
22. Objective Turret
23. Indicating Desiccant Cartridge
24. Rubber Plug
ASSEMBLY PROCEDURE
The Universal Mount (Item 18) may be secured to the bench by
using either the C Clamp (Item 20) or by using three screws
located in the counter bored holes of the Mount. The Mount is
simply a cylindrical base plate, with a vertical stanchion in
the center, upon which the column base is placed. You either
secure the mount to an edge of your bench with the C clamp, or
screw the mount to table at any location. Note: The Mount may
be used as a template for the hole positions.
Place the Column Base (Item 19) on top of the Universal
Mount (Item 18) then carefully lower the Counter Balanced
Stand (Item 15) onto the spigot.
Lightly tighten the Height Friction Clamp (Item 6) and
remove the Nut and Washer (Item 2) from the Yaw Spigot (Item
5).
Remove Cover Plate (Item 1) from Head and carefully lower
the Mantis over the Yaw Spigot (Item 5).
Replace the Washer and Nut (Item 2), do not lock solid, Head
should rotate on the Yaw Spigot (Item 5) with the required
degree of friction replace Cover Plate (Item 1).
Screw the chosen Objectives onto the Turret (Item 22).
Connect the low Voltage Plug (Item 12) into the socket on
the rear of the head (Item 7).
Check that the Voltage Selector (Item 17) is correctly set
then connect to the main supply. The instrument is now ready
for use. A regular check should be made to ensure the Cooling
Fan is running.
OPERATION
Select the required magnification using the Objective
Selection Lever located on left hand side of Head (Item 9).
Position the Viewing Head at a comfortable height and adjust
Friction Clamp (Item 6) as required.
Turn on the Supply Switch (Item 14) and hold the subject
beneath the Objectives in the optimum position for viewing.
Focusing can be carried out by raising the subject (it hand
held) or by moving the Mantis Head up or down.
Adjust the Eye Spacing Control (Item 3) to obtain a
comfortable stereo view. Note: That incorrect adjustment can
result in inverted stereo. The aim is to obtain good stereo
with max head freedom.
The angle of lighting can be altered by rotation of Edge
Wheel (Item 11).